3
•
Detector Overview
Page 32
Vanquish Charged Aerosol Detectors Operating Manual
The largest aerosol droplets condense within the spray chamber (no. 5)
and are removed by the drain pump (no. 6) through the waste outlet.
Smaller droplets pass into a temperature-controlled evaporation tube
(no. 7), in which solvent and more volatile components are vaporized.
By the end of the evaporation tube, the resulting aerosol particles enter
the mixing chamber (no. 8), where they collide with a secondary stream
of ionized gas (no. 9) that has passed through a corona charger (no. 10).
Each dried aerosol particle undergoes surface diffusional charging,
where the level of charge per particle is proportional to particle size and
the cumulative charge of the population is directly related to the analyte
amount.
In the ion trap (no. 11), higher mobility ions and smaller charged
particles are removed. Lower mobility charged particles pass to a
conductive filter and aggregate charge is measured with a sensitive
electrometer (no. 12). Signal is processed by the electrometer board
(no. 13) and sent to the chromatography data system. The resulting gas
flow exits the detector through the gas exhaust (no. 14).
Содержание Vanquish VF-D20
Страница 4: ...Contacting Us Page 4 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 28: ...2 Safety Page 28 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 38: ...3 Detector Overview Page 38 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 44: ...4 Unpacking and Transport Page 44 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 150: ...7 Maintenance and Service Page 150 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 177: ...12 Index Vanquish Charged Aerosol Detectors Operating Manual Page 177 12 Index...
Страница 181: ...12 Index Vanquish Charged Aerosol Detectors Operating Manual Page 181...
Страница 182: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...