Vanquish Charged Aerosol Detectors Operating Manual
5.6
Setting Up the Liquid Flow Connections
5.6.1
General Information and Guidelines
This section describes how to set up the flow connections to and from
the detector and additional flow connections if required.
When setting up flow connections, follow these rules and
recommendations:
Flow connections can be filled with hazardous substances. Observe the
warning messages and precautionary statements presented in
section 2.3 Safety Precautions
page 19
NOTICE
Before
you connect the charged aerosol detector in the system
flow path:
When you install devices or components to the system, always flush
them to waste before connecting them in the system flow path. To flush
the Vanquish modules, follow the instructions in the
Vanquish System
Operating Manual
.
Flushing the charged aerosol detector will be performed later during
equilibration.
•
Dirty components can contaminate the chromatography system.
Contamination leads to poor performance of the modules and entire
system or can even cause damage to the modules and system.
Therefore:
♦
Always wear appropriate gloves.
♦
Place the components only on a clean, lint-free surface.
♦
Keep your tools clean.
♦
Use only lint-free cloth for cleaning.
•
For installation instructions and guidelines and for handling
recommendations, see
section 5.6.3 Connecting Fittings, Capillaries,
and Tubing, page 73
Содержание Vanquish VF-D20
Страница 4: ...Contacting Us Page 4 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 28: ...2 Safety Page 28 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 38: ...3 Detector Overview Page 38 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 44: ...4 Unpacking and Transport Page 44 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 150: ...7 Maintenance and Service Page 150 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 177: ...12 Index Vanquish Charged Aerosol Detectors Operating Manual Page 177 12 Index...
Страница 181: ...12 Index Vanquish Charged Aerosol Detectors Operating Manual Page 181...
Страница 182: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...