7
•
Maintenance and Service
Vanquish Charged Aerosol Detectors Operating Manual
Page 135
♦
The top gas filter must be the HEPA gas filter (thin white inner
filter cylinder; no. 1 in
7. Push the new gas filter assembly into the clamps observing the
orientation signs for the gas flow on the gas filters. See
Figure 35: Correct orientation of the gas filter assembly
No.
Description
1
HEPA gas filter
2
Charcoal gas filter
8. Push the (push-in) elbow fittings into the gas connectors on the
detector. Check that the connections are secure: Slightly pull on the
fittings. If a fitting slides out, repeat this step.
9. Install the protective cover: Hold the protective cover onto the
interior panel. Slide it downward carefully so that the bottom screws
match the recesses.
10. With the screwdriver, tighten the three screws on the protective
cover.
11. Turn on the detector with its main power switch.
12. In Chromeleon, consider updating the Predictive Performance
information for the gas filter assembly.
13. Turn on the gas flow. See
section 6.5.3 Turning On the Detector Gas
Содержание Vanquish VF-D20
Страница 4: ...Contacting Us Page 4 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 28: ...2 Safety Page 28 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 38: ...3 Detector Overview Page 38 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 44: ...4 Unpacking and Transport Page 44 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 150: ...7 Maintenance and Service Page 150 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 177: ...12 Index Vanquish Charged Aerosol Detectors Operating Manual Page 177 12 Index...
Страница 181: ...12 Index Vanquish Charged Aerosol Detectors Operating Manual Page 181...
Страница 182: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...