5
•
Installation
Page 68
Vanquish Charged Aerosol Detectors Operating Manual
2. Connect the gas exhaust tubing to the ventilation and route it to the
detector.
3. Check the length of the gas tubes.
The gas tubes should go straight to the detector. If the gas tubes are
longer than required, use a tubing cutter to cut the tubes straight to
the required length. Make sure that the cut is at a right angle to the
length of the gas tubes.
4. Connect the gas tubes to the elbow connectors. Mind the correct
orientation of the elbow connectors (see
a) Connect the gas inlet tubing to the gas inlet elbow connector.
b) Connect the gas exhaust tubing to the gas exhaust elbow
connector.
Figure 17: Connecting gas tubes to the elbow connectors
No. Description
No. Description
1
Connection stem to gas inlet port 4
Connection stem to gas exhaust port
2
Connector for gas inlet tubing
5
Connector for gas exhaust tubing
3
Gas inlet tubing
6
Gas exhaust tubing
Содержание Vanquish VF-D20
Страница 4: ...Contacting Us Page 4 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 28: ...2 Safety Page 28 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 38: ...3 Detector Overview Page 38 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 44: ...4 Unpacking and Transport Page 44 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 150: ...7 Maintenance and Service Page 150 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 177: ...12 Index Vanquish Charged Aerosol Detectors Operating Manual Page 177 12 Index...
Страница 181: ...12 Index Vanquish Charged Aerosol Detectors Operating Manual Page 181...
Страница 182: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...