6
•
Operation
Page 108
Vanquish Charged Aerosol Detectors Operating Manual
Guidelines for selecting the evaporation temperature
The following guidelines are recommended for selecting an optimal
evaporation temperature:
•
In general, use the lowest evaporation temperature within the
specified temperature range that provides acceptable limits of
sensitivity.
•
As a starting point, use an evaporation temperature of 35 °C.
•
Higher evaporation temperature settings can be used when analytes
of interest have low volatility, relative to the background. Since the
volatility of background residue is typically unknown,
experimentation is required for optimizing this parameter for a given
method.
•
Perform screening runs at two evaporation temperature settings.
♦
With the VH-detector, use a minimum of two evaporation
temperatures. Typically, 35 °C and 40 °C are appropriate if the
temperature range is unknown. If required, perform additional
tests in a maximum of 5 °C increments.
♦
With the VF-detector, use two evaporation temperatures,
depending on whether the temperature range is known. If the
temperature range is unknown, 35 °C and 50 °C is an
appropriate starting point for screening runs.
•
Consider that higher evaporation temperature settings may
inadvertently reduce response for low-analyte levels, even if they
are considered as a non-volatile analytes.
•
Remember to check the limits of sensitivity by actual analysis of low-
level standards. Also, check the analyte signal reproducibility at each
evaporation temperature setting.
•
For detection of semi-volatile analytes, you can use lower
evaporation temperature settings. However, this may lead to higher
background levels and noise.
NOTICE
Do not use non-volatile modifiers, such as sodium phosphate.
Mobile Phase Guidelines for the Detector in section 6.5.1,
page 95
Содержание Vanquish VF-D20
Страница 4: ...Contacting Us Page 4 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 28: ...2 Safety Page 28 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 38: ...3 Detector Overview Page 38 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 44: ...4 Unpacking and Transport Page 44 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 150: ...7 Maintenance and Service Page 150 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 177: ...12 Index Vanquish Charged Aerosol Detectors Operating Manual Page 177 12 Index...
Страница 181: ...12 Index Vanquish Charged Aerosol Detectors Operating Manual Page 181...
Страница 182: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...