CAUTION—Hydrostatic Pressure
Solvent may spill when you open the flow path during pump operation.
This is due to hydrostatic pressure in the system when the solvent
reservoirs are located above the pump outlet.
Before you loosen a connection in the low-pressure flow path, turn off
the pump flow and wait until the system pressure is down to zero.
Unscrew the caps of the solvent reservoirs and remove the solvent lines
together with the caps from the reservoirs. Empty the solvents lines.
Retighten the reservoir caps.
NOTICE—Detector Gas Flow and Pump Flow
Improper handling of detector gas flow and pump flow start and stop
can damage the detector. Observe the respective order of steps:
•
When turning on detector gas flow and pump flow
Turn on the detector gas flow first, wait at least 5 minutes and only
then turn on the pump flow.
•
When turning off pump flow and detector gas flow
Turn off the pump flow first, wait at least 5 minutes and only then
turn off the detector gas flow. During the waiting period, the drain
pump drains any remaining mobile phase to waste before it
automatically turns off.
Содержание Vanquish VF-D20
Страница 4: ...Contacting Us Page 4 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 28: ...2 Safety Page 28 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 38: ...3 Detector Overview Page 38 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 44: ...4 Unpacking and Transport Page 44 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 150: ...7 Maintenance and Service Page 150 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 177: ...12 Index Vanquish Charged Aerosol Detectors Operating Manual Page 177 12 Index...
Страница 181: ...12 Index Vanquish Charged Aerosol Detectors Operating Manual Page 181...
Страница 182: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...