7
•
Maintenance and Service
Vanquish Charged Aerosol Detectors Operating Manual
Page 149
d) Push the slide-in module onto the rails and into the enclosure
until the slide-in module sits completely in the enclosure.
Figure 40: Inserting the slide-in module
4. Tighten the four captive screws on the slide-in module hand-tight.
NOTICE
Verify that the screws are tightened. Pull the slide-in module
by the leak tray towards the front and check whether the screws
move. If they do not move, the slide-in module is installed properly.
If the screws move, tighten the screws further.
5. Set up the slide-in module:
a) To connect the slide-in module and to set up flow connections,
follow the instructions in the
Installation chapter
in this
operating manual.
b) To prepare the slide-in module for first-time operation, follow
the instructions in the
Preparing the Detector for Operation
section
in this operating manual.
6. Prepare
all other
modules of the Vanquish system for operation and
restart them. Refer to the
Operating Manuals
for the modules.
7. Before starting an analysis, let the chromatography system
equilibrate and be sure that it is ready for operation.
Содержание Vanquish VF-D20
Страница 4: ...Contacting Us Page 4 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 28: ...2 Safety Page 28 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 38: ...3 Detector Overview Page 38 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 44: ...4 Unpacking and Transport Page 44 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 150: ...7 Maintenance and Service Page 150 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 177: ...12 Index Vanquish Charged Aerosol Detectors Operating Manual Page 177 12 Index...
Страница 181: ...12 Index Vanquish Charged Aerosol Detectors Operating Manual Page 181...
Страница 182: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...