7
•
Maintenance and Service
Vanquish Charged Aerosol Detectors Operating Manual
Page 147
4. Remove the slide-in module from the enclosure.
The following steps require a team effort:
a) Take the slide-in module on both sides from below.
b) Pull the slide-in module from the rails towards the front.
c) Place the slide-in module on a clean and stable surface.
5. Return the slide-in module:
a) To request a dedicated packaging for the slide-in module and for
the appropriate procedure for returning the module, contact
your local Thermo Fisher Scientific support organization.
b) Pack the slide-in module in the dedicated packaging.
NOTICE
The packaging for the slide-in module differs from the
packaging for the complete module. Shipping the slide-in module or
the detector incorrectly leads to damage on the device. Always pack
and ship the slide-in module and the detector in the respective,
dedicated packaging.
CAUTION—Possible Contamination
Hazardous substances may have contaminated the device during
operation and may cause personal injury to service personnel.
•
Decontaminate all parts of the device that you want to return for
repair.
•
Fill in and sign the Decontamination Certificate, which is part of the
Service Return Form. Sign the certificate to confirm that the device
has been properly decontaminated and that it is free of hazardous
substances.
•
Thermo Fisher Scientific refuses to accept devices for repair if the
Decontamination Certificate is missing.
Содержание Vanquish VF-D20
Страница 4: ...Contacting Us Page 4 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 28: ...2 Safety Page 28 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 38: ...3 Detector Overview Page 38 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 44: ...4 Unpacking and Transport Page 44 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 150: ...7 Maintenance and Service Page 150 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 177: ...12 Index Vanquish Charged Aerosol Detectors Operating Manual Page 177 12 Index...
Страница 181: ...12 Index Vanquish Charged Aerosol Detectors Operating Manual Page 181...
Страница 182: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...