Maintenance and Service
Page 130
Vanquish Charged Aerosol Detectors Operating Manual
2. Pull the nebulizer straight out of the opening. Avoid the nebulizer tip
from touching the sides of the nebulizer opening, as this can damage
the tip.
Figure 28: Removing the nebulizer
3. Install the dust seal plug to the opening for the nebulizer.
4.
If the nebulizer is to be transported
Pack the nebulizer:
a) Carefully install the protective tubing to the nebulizer spray
pipe. Do not to touch the nebulizer tip with the protective
tubing, as it can damage the tip.
b) Install the plug to the nebulizer inlet.
Figure 29: Nebulizer with protective tubing and inlet plug
c) Pack the nebulizer in its original packaging that it was shipped in.
To re-install the nebulizer, follow the instructions in
section 5.5.3,
page 62