HistoStar Operator Guide A81010100 Issue 2
86
Cold Plate
Problem
Possible Cause
Remedy
Inadequate cooling
Ambient temperature
is above 27°C
Reduce the ambient temperature or increase air circulation
around instrument
Condenser air-flow is
blocked
Remove blockage from fan inlets
Instrument
malfunction
Call Thermo Fisher Scientific Service Department
Compressor will not
switch on
Cold module not
powered
Check fuses and replace if necessary - see Fuse Replacement
Check that Mains lead is connected correctly
Cold module set for
Timer operation but
not connected to
embedding module
Either:
Connect to embedding module - see Connections
or
Change setting to 'Compressor On'
Power not switched on
via embedding module
Switch cold plate on via cooling options screen - see
Temperature Set Points