Chapter 2 - Basic Operation
23
•
Transfer the base mould and cassette onto the cold plate for rapid chilling.
•
Once fully cooled and set, the cassette can be removed from the base mould ready for sectioning.
Notes
A frost build-up on the cold plate is normal. The rate of frost build-up will vary with the humidity in the
laboratory.
Frost build-up will not normally interfere with cooling, but if excessive frost develops turn the Cold Module off for
a short period until defrosting occurs.
Remove any water from the cold plate surface using a dry cloth.
An hour on the cold plate should be more than adequate to ensure that the wax is fully set. Prolonged chilling can
cause minute cracks to form in the wax at the bottom of the base mould.