HistoStar Operator Guide A81010100 Issue 2
38
Note
There is no need to leave a gap between individual modules.
•
The front edge of the HistoStar modules should be level with the front edge of the work-bench as shown.
•
HistoStar must be located within easy reach of a suitably rated power supply - see System Specification.
•
The work-bench must be capable of safely supporting all the HistoStar modules, including the weight of the
wax and accessories - see System Specification.
•
The work-bench must be level and made of non-flammable material.
•
Do not position the HistoStar directly under air-conditioning or heating vents, or near to an open window.
•
Do not position the HistoStar in an area where it is likely that the operator could be bumped or knocked
during embedding.
•
Keep out of direct sunlight.
Levelling
To ensure proper wax drainage during use, the HistoStar must be levelled correctly.
To level the HistoStar :
•
Use a spanner to loosen the lock-nut on each of the adjustable feet.