Chapter 4 - Troubleshooting
85
Cold Spot
Problem
Possible Cause
Remedy
Inadequate cooling
Surface of cold spot has
not reached the required
temperature yet
Allow sufficient time for the cold spot to cool.
Fan air flow blocked
Remove blockage from fan inlet directly underneath cold spot
Instrument malfunction Switch the embedding module off for 30 seconds and then
back on again to reset the control system
Call Thermo Fisher Scientific Service Department