Chapter 4 - Troubleshooting
81
Hot Spot
Problem
Possible Cause
Remedy
Hot spot will not heat up Hot spot heater not
switched on
Switch on the hot spot heater - see Temperature Set Points
Instrument malfunction Switch the embedding module off for 30 seconds and then
back on again to reset the control system
Compare the set point temperature to the actual temperature
- if the set point temperature is not achieved after a suitable
length of time:
•
Switch off the embedding module
•
Wait for 20 minutes and then switch the embedding
module back on
Perform Heater System Check
Call Thermo Fisher Scientific Service Department
Hot spot lights do not
switch on
Lighting not switched
on
Switch on the hot spot lights - see Instrument Lighting
Instrument malfunction Switch the embedding module off for 30 seconds and then
back on again to reset the control system
Perform Heater System Check
Call Thermo Fisher Scientific Service Department