HistoStar Operator Guide A81010100 Issue 2
80
Forceps Block
Problem
Possible Cause
Remedy
Does not heat up
sufficiently to melt wax off
forceps
Power level too low
for wax in use
Increase the power level of the forcep block - see Temperature
Set Points
Does not heat up
Instrument
malfunction
Switch the embedding module off for 30 seconds and then
back on again to reset the control system
Perform Heater System Check
Call Thermo Fisher Scientific Service Department