Chapter 4 - Troubleshooting
79
Wax tank will not heat
up
Wax tank heater not
switched on
Switch on the wax tank heater - see Temperature Set Points
Instrument
malfunction
Switch the embedding module off for 30 seconds and then back on
again to reset the control system
Compare the set point temperature to the actual temperature - if
the set point temperature is not achieved after a suitable length of
time:
•
Switch off the embedding module
•
Wait for 20 minutes and then switch the embedding module
back on
Perform Heater System Check
Call Thermo Fisher Scientific Service Department
Excessive wax drips
from the dispenser
Debris blocking the
valve seal
Call Thermo Fisher Scientific Service Department
Wax dispenses
continuously
Dispense lever
jammed
Check operation of dispense lever
Call Thermo Fisher Scientific Service Department
Footswitch activated Check that footswitch is not being depressed accidentally
Check footswitch tubing for kinks