Tissue Storage Tank
Problem
Possible Cause
Remedy
Wax not fully melted
Wax has not yet
melted
Allow time for wax to melt
Ensure temperature setting is high enough to melt the wax
being used
Compare the set point temperature to the actual temperature -
if the set point temperature is not achieved after a suitable
length of time:
•
Switch off the embedding module
•
Wait for 20 minutes and then switch the embedding
module back on
If problem persists call Thermo Fisher Scientific Service
Department
Wax surface not
melted
Close tissue storage cover during warm-up as draughts can
cause a skin to form on the surface of the wax.
Tank does not heat up
Instrument
malfunction
Switch the embedding module off for 30 seconds and then
back on again to reset the control system
Perform Heater System Check
Call Thermo Fisher Scientific Service Department