2 • Description
Doc. 155028-01 08/20
19
Pulse Damper
Flow output from the pressure transducer continues to the pulse damper,
which smooths minor pressure variations. From there, flow is directed to
the injection valve and then to the remainder of the chromatography
system.
2.3.3
Injection Valve with Sample Loop
The injection valve is a six-port, electrically-activated Rheodyne valve.
The valve has two operating positions: Load and Inject. Eluent flows
through either the Load or Inject path, depending on the valve position. A
10
L sample loop (P/N 042949) is installed on the valve at the factory.
shows flow schematics for the valve.
•
In the Load position, sample is loaded into the sample loop, where it
is held until injection. Eluent flows from the pump, through the valve,
and to the column, bypassing the sample loop. Sample flows from the
syringe (or the Dionex AS-DV sample out line, if installed), through
the valve, and into the sample loop. Excess sample flows out to waste.
•
In the Inject position, sample is swept to the column for analysis.
Eluent flows from the pump, through the sample loop, and on to the
Figure 2-6. Injection Valve Flow Schematics
LOAD POSITION
INJECT POSITION
= Sample
= Eluent
L
P
C
To Waste
Sample In
To Column
From Pump
Sample In
Sample Loop
To Column
To Waste
From Pump
L
S
W
Содержание Dionex Easion
Страница 1: ...Dionex Easion Ion Chromatography System Operator s Manual Document No 155028 Revision 01 August 2020 ...
Страница 34: ...Dionex Easion Operator s Manual 26 Doc 155028 01 08 20 ...
Страница 68: ...Dionex Easion Operator s Manual 60 Doc 155028 01 08 20 ...
Страница 76: ...Dionex Easion Operator s Manual 68 Doc 155028 01 08 20 Figure 5 2 Dionex Easion Flow Schematic ...
Страница 110: ...Dionex Easion Operator s Manual 102 Doc 155028 01 08 20 ...
Страница 118: ...Dionex Easion Operator s Manual 110 Doc 155028 01 08 20 ...