D • FAQ
Doc. 155028-01 08/20
101
D.9
How do I clean a column?
Refer to the troubleshooting section of the column manual.
D.10 Why is the conductivity high?
Possible reasons for high conductivity include:
•
The suppressor needs regeneration. Refer to the troubleshooting section of the
suppressor manual.
•
The regenerant is exhausted and should be remade. See
for
instructions for how to prepare regenerant.
•
The cell is out of calibration. Contact Technical Support for Dionex products
for assistance.