Dionex Easion Operator’s Manual
12
Doc. 155028-01 08/20
Conductivity Cell
The flow-through heated conductivity cell measures the electrical
conductance of analyte ions as they pass through the cell. A heat
exchanger inside the cell regulates the temperature to 40 °C (104 °F). The
cell is housed inside a Thermo Scientific
™
Dionex
™
DS5 Detection
Stabilizer (P/N 067761). For more information about the conductivity cell
and Dionex DS5 Detection Stabilizer, see
.
Suppressor
The suppressor reduces the eluent conductivity and enhances the
conductivity of the sample ions, thereby increasing detection sensitivity.
The Thermo Scientific
™
Dionex
™
Chemically Regenerated Suppressor
(Dionex CRS
™
500) is recommended for use with the Dionex Easion. For
more information about suppressors, see
.
Separator and Guard Columns
Both the separator and guard columns are packed with resin and perform
the separation of the sample ions. The main function of the guard column
is to trap contaminants and particulates that might damage the separator
column.
Pressure Transducer
The pressure transducer measures the system backpressure. For more
information about the pressure transducer, see
Pump Heads
The Dionex Easion includes a dual-piston serial pump. The flow rate can
be set from 0.01 mL/min to 5.00 mL/min. However, for optimum
performance, set the flow rate to between 0.20 and 3.00 mL/min. Setting
the flow rate to 0.00 mL/min turns off the pump. For more information
about the pump, see
.
Injection Valve
The injection valve is a six-port, electrically-activated valve. For more
information about the injection valve, see
.
Содержание Dionex Easion
Страница 1: ...Dionex Easion Ion Chromatography System Operator s Manual Document No 155028 Revision 01 August 2020 ...
Страница 34: ...Dionex Easion Operator s Manual 26 Doc 155028 01 08 20 ...
Страница 68: ...Dionex Easion Operator s Manual 60 Doc 155028 01 08 20 ...
Страница 76: ...Dionex Easion Operator s Manual 68 Doc 155028 01 08 20 Figure 5 2 Dionex Easion Flow Schematic ...
Страница 110: ...Dionex Easion Operator s Manual 102 Doc 155028 01 08 20 ...
Страница 118: ...Dionex Easion Operator s Manual 110 Doc 155028 01 08 20 ...