Doc. 155028-01 08/20
61
5 • Service
This chapter describes Dionex Easion service and repair procedures that users can
perform. Procedures not included here, including electronics-related repair
procedures, must be performed by Thermo Fisher Scientific personnel.
For assistance, contact Technical Support for Dionex products before proceeding:
•
In the U.S. and Canada, call 1-800-532-4752 and select
option 2
.
•
Outside the U.S. and Canada, call the nearest Thermo Fisher Scientific office.
Before replacing any part, refer to the troubleshooting information in
correctly identify the cause of the problem.
5.1
Diagnostics and Calibrations
The Wellness panel in Chromeleon provides controls for performing diagnostic
and calibration functions for the Dionex Easion.This section provides an overview
of Wellness panel functions.
Use Wellness panel controls to perform the following procedures:
•
Calibration of the pressure transducer (see
•
Calibration of the flow rate (see
Substituting non-Thermo Scientific/Dionex parts may impair system
performance, thereby voiding the product warranty. For details, see
the warranty statement in the Thermo Scientific/Dionex Terms and
Conditions.
Содержание Dionex Easion
Страница 1: ...Dionex Easion Ion Chromatography System Operator s Manual Document No 155028 Revision 01 August 2020 ...
Страница 34: ...Dionex Easion Operator s Manual 26 Doc 155028 01 08 20 ...
Страница 68: ...Dionex Easion Operator s Manual 60 Doc 155028 01 08 20 ...
Страница 76: ...Dionex Easion Operator s Manual 68 Doc 155028 01 08 20 Figure 5 2 Dionex Easion Flow Schematic ...
Страница 110: ...Dionex Easion Operator s Manual 102 Doc 155028 01 08 20 ...
Страница 118: ...Dionex Easion Operator s Manual 110 Doc 155028 01 08 20 ...