5 • Service
Doc. 155028-01 08/20
69
5.4
Cleaning Eluent Bottles
Before preparing new eluent, all eluent bottles should be rinsed thoroughly (inside
and out) with ASTM Type I (18.2 megohm-cm) filtered and deionized water. If a
bottle still appears dirty, or if there is a slimy film on the inside, clean the bottle as
instructed below.
1. Dispose of any remaining chemicals according to municipal regulations.
2. Rinse the bottle (inside and out) with ASTM Type I (18.2 megohm-cm)
filtered and deionized water.
3. Rinse the inside of the bottle with isopropyl alcohol or methanol.
4. If algae or bacteria have left a slimy film on the bottle, use an algicide or
disinfectant (for example, dilute hydrogen peroxide).
5. Rinse cleaning chemicals out of the bottle with ASTM Type I (18.2 megohm-
cm) filtered and deionized water.
6. Dry the bottle with clean, particulate-free air.
5.5
Changing the Sample Loop
A 10
L PEEK sample loop (P/N 042949) is installed on the injection valve at the
factory. Unless you are using a 2 mm column, you can replace the sample loop
with another 10
L PEEK sample loop. When using a 2 mm column, refer to the
column manual for the loop size required.
1. Click the pump
Off
button
on the Chromeleon ePanel.
2. Open the Dionex Easion
front door.
3. Disconnect the sample loop
from ports
L
and
L
on the
Figure 5-3. Dionex Easion Injection Valve
Sample Loop
Содержание Dionex Easion
Страница 1: ...Dionex Easion Ion Chromatography System Operator s Manual Document No 155028 Revision 01 August 2020 ...
Страница 34: ...Dionex Easion Operator s Manual 26 Doc 155028 01 08 20 ...
Страница 68: ...Dionex Easion Operator s Manual 60 Doc 155028 01 08 20 ...
Страница 76: ...Dionex Easion Operator s Manual 68 Doc 155028 01 08 20 Figure 5 2 Dionex Easion Flow Schematic ...
Страница 110: ...Dionex Easion Operator s Manual 102 Doc 155028 01 08 20 ...
Страница 118: ...Dionex Easion Operator s Manual 110 Doc 155028 01 08 20 ...