Density Calculation
Slurry Material
Thermo Scientific
DensityPRO User Manual
C-11
Pounds per cubic yard (lb/yd3)
Kilogram per cubic meter (kg/m3)
ID: Pipe ID – Inside diameter of pipe in units selected by the user.
Pipe ID Units: Units of measure for the inside diameter of the pipe.
Centimeters (cm)
Millimeters (mm)
Inches (in)
Feet (ft)
Yards (yd)
Meters (m)
Dc: Density of the carrier material – Density of the carrier in grams per cubic
centimeter.
µc: Attenuation of the carrier material – The value of the attenuation
coefficient depends on the composition of the process material and the type of
source used, cesium-137 (Cs-137) or cobalt-60 (Co-60).
Ds: Density of the solid material – Density of the solids in grams per cubic
centimeter.
(µs): Attenuation coefficient of the solids – The value of the attenuation
coefficient depends on the composition of the process material and the type of
source used, cesium-137 (Cs-137) or cobalt-60 (Co-60).
(Ibkg): Background radiation – Background radiation at the detector in counts
per second (CPS).
Calibration density units – The user can select the density units used for
calibration from the following options.
Grams per cubic centimeter (g/cm3)
Pounds per US gallon (lb/gal)
Pounds per Imperial gallon (lb/ImpGal)
Pounds per cubic foot (lb/ft3)
Short tons per cubic yard (STon/yd3)
Long tons per cubic yard (LTon/yd3)
Grams per liter (g/L)
Ounces per cubic inch (oz/in3)
Pounds per cubic inch (lb/in3)
Содержание DensityPRO NAI Series
Страница 2: ......
Страница 4: ......
Страница 6: ...This page intentionally left blank ...
Страница 8: ...This page intentionally left blank ...
Страница 14: ...xiv DensityPRO User Manual Thermo Scientific This page intentionally left blank ...
Страница 22: ...A xxii DensityPRO User Manual Thermo Scientific Quick Setup This page intentionally left blank ...
Страница 124: ...Operation Application Screens 4 72 DensityPRO User Manual Thermo Scientific Figure 4 39 High and Low Alarms ...
Страница 134: ......
Страница 158: ......
Страница 195: ......