Operation
Application Screens
Thermo Scientific
DensityPRO User Manual
4-59
To access the information on Application Standardization tabs using only the remote
transmitter keypad, see
Figure 4-29.
Application Screen, Standardization Tab
1.
Select a condition from the Standardization On dropdown.
None
Pipe Empty
Pipe Full
Bypassed
2.
Standardization counts are averaged over a user-defined period of time. Enter a
time period, in seconds, in the STD Sample Time (sec) textbox. The time can
range from 1 second to 65,535 seconds.
3.
When the Standardization On field is set as Pipe Full, the STD Density Value
textbox becomes available to the user. Enter a standardization density value. The
units displayed are those associated with the primary density measurement.
4.
The following commands are available for the standardization:
Start STD – Begin the standardization cycle.
Abort STD – Abort or reject the standardization cycle before cycle
completion. When the user terminates the standardization cycle, the value of
the standardization counts and the standardization date and time stored in the
gauge remain unchanged.
Содержание DensityPRO NAI Series
Страница 2: ......
Страница 4: ......
Страница 6: ...This page intentionally left blank ...
Страница 8: ...This page intentionally left blank ...
Страница 14: ...xiv DensityPRO User Manual Thermo Scientific This page intentionally left blank ...
Страница 22: ...A xxii DensityPRO User Manual Thermo Scientific Quick Setup This page intentionally left blank ...
Страница 124: ...Operation Application Screens 4 72 DensityPRO User Manual Thermo Scientific Figure 4 39 High and Low Alarms ...
Страница 134: ......
Страница 158: ......
Страница 195: ......