Startup & the Setup Wizard
EZ Cal II Control Buttons
3
-
4
DensityPRO User Manual
Thermo Scientific
4.
In order to save the configuration to the gauge permanently, click the Write to
Flash Again button. This will ensure that the specified parameters are loaded each
time the system boots.
Sometimes it may be preferable to load parameters into the gauge for only a
short time, such as while running a specific test. In this instance, once the
parameters have been downloaded to the gauge, the process is complete. Once
the gauge is shut down, the configuration parameters will revert to the last
configuration written to flash.
EZ Cal II Control
Buttons
The following buttons are seen consistently throughout the EZ Cal II software:
–
Abort – Stops the current process.
–
Auto Refresh – Continuously refreshes the screen with data from the gauge.
–
Back – Moves the user to the previous screen.
–
Cancel – Closes the current screen.
–
Home – Returns the user to the Detector and Application Type Selection screen,
–
Next – Moves the user to the next available screen.
–
Refresh – Updates the screen with the latest information from the gauge.
–
Submit – Saves any information entered on the current screen to the gauge.
If a screen has multiple Submit buttons, each button will apply to a separate,
portioned-off area of the screen.
User Modes &
Passwords
Password protection limits access to the gauge’s setup parameters based on the access
level rights granted through the four user modes below.
–
Display List –This is normal mode of operation. The local display will scroll
through a list of measurement data with no ability to view or modify the database
via Remote Backplane (RPB) display/keypad. Remote read access to the database is
available however no writes to are allowed.
–
Operator Mode – The local display on the RBP will be running the user menu
interface, allowing a user to view the configuration of the gauge. Remote read
access to the database is available, however no writes to are allowed.
–
Technician Mode – The local user can configure a small sub-set of the database
within the gauge via the front panel menu system on the RBP. Remote write access
is also available to a small subset of the gauge database so that calibration and
standardization can be performed.
Содержание DensityPRO NAI Series
Страница 2: ......
Страница 4: ......
Страница 6: ...This page intentionally left blank ...
Страница 8: ...This page intentionally left blank ...
Страница 14: ...xiv DensityPRO User Manual Thermo Scientific This page intentionally left blank ...
Страница 22: ...A xxii DensityPRO User Manual Thermo Scientific Quick Setup This page intentionally left blank ...
Страница 124: ...Operation Application Screens 4 72 DensityPRO User Manual Thermo Scientific Figure 4 39 High and Low Alarms ...
Страница 134: ......
Страница 158: ......
Страница 195: ......