Flashing the Application Firmware
Flash Application Firmware
Thermo Scientific
DensityPRO User Manual
B-3
Figure B-3.
Flash Application Screen
6.
Select the firmware to flash and navigate to the correct file.
7.
Choose the correct S19 file and click the Open button.
Figure B-4.
Open S19 File
8.
Cycle the system power.
9.
From the Flash Application Screen, click the Start Flash button to start
programming the Main CPU.
10.
Once the application has flashed successfully, close EZ Cal II.
11.
Cycle the system power. The system is now ready for programming.