Quick Setup
Calibration
x
x
i
DensityPRO User Manual
Thermo Scientific
Calibration
There are two types of calibration: a one point calibration and a two point calibration.
The two calibration points allow you to have two calibration points in the region of
interest, and standardization can then be used to compensate for pipe wear and process
buildup on the pipe walls. The first point calibration can be thought of as an offset. It
will move the response curve up or down.
The second point calibration is a slope correction. The slope correction pivots around
the first cal point. The slope can be set using the second point calibration or entered
directly.
If the gauge is standardized on water or on process, the standardization serves as the
first calibration point.