Operation
Calibration
4
-
80
DensityPRO User Manual
Thermo Scientific
b.
To recalculate the point, click Start Calibrate before clicking the Submit
button.
c.
If the point data is satisfactory, click Submit to save and Next to calibrate the
next point.
In a 2-point calibration, the next screen will calibrate the maximum
values.
In a 3-point calibration, the next screen will calibrate the mid-point
values. Submit that information and click the Next button to proceed to
the maximum point value screen.
7.
Follow steps 4 – 6 to calibrate the remaining point(s).
Figure 4-44.
Output Calibration, Maximum Point
8.
Once the CAL Max Point data has been submitted, click the Next button to reach
accept or reject the calibration on the CAL Point Completed Screen.