Operation
Application Screens
4
-
60
DensityPRO User Manual
Thermo Scientific
Accept STD – Accept the standardization value before or after the cycle
completes. When the user accepts the standardization value before the cycle
completes, the date, time and value of the standardization counts at the time
of acceptance are stored in the gauge.
Reject STD – Reject the standardization value after completion of the
standardization cycle. If the user rejects the standardization, the value of the
standardization counts and the standardization date and time stored in the
gauge are unchanged.
Note:
The high voltage must be stable during the standardization cycle. If the
high voltage is unstable, the cycle will abort.
Gauge Calibration
When a calibration is required, a one point calibration provides a reference
measurement at one density in the range of interest. This form of calibration is
sufficient in most instances. If greater measurement accuracy is required, two-point
calibration can be performed. The calibration density value is currently entered only in
g/cc. In future software releases the user will be able to select the calibration density
units.
To access the information on Application Gauge Calibration tabs using only the
remote transmitter keypad, see
Figure 4-30.
Application Screen, Gauge Calibration CAL Data Tab
Содержание DensityPRO NAI Series
Страница 2: ......
Страница 4: ......
Страница 6: ...This page intentionally left blank ...
Страница 8: ...This page intentionally left blank ...
Страница 14: ...xiv DensityPRO User Manual Thermo Scientific This page intentionally left blank ...
Страница 22: ...A xxii DensityPRO User Manual Thermo Scientific Quick Setup This page intentionally left blank ...
Страница 124: ...Operation Application Screens 4 72 DensityPRO User Manual Thermo Scientific Figure 4 39 High and Low Alarms ...
Страница 134: ......
Страница 158: ......
Страница 195: ......