3.
In the Min Value @ Min Volt textbox, enter the minimum value, in engineering
units, that corresponds to the value entered in the Minimum Volt textbox.
4.
In the Max Value @ Max Volt textbox, enter the maximum value, in engineering
units, that corresponds to the value entered in the Maximum Volt textbox.
5.
Additional fields on this screen, which are listed below, are for display purposes
only and are not configurable by the user.
Live Value (EU) – This field displays the live calibrated value in engineering
units in both Live and Manual modes.
Raw Value (Volt) – This field displays the live calibrated value in volts in both
Live and Manual modes.
#1 Volt Min Value – This field displays the expected minimum voltage input
value provided by the user during calibration.
#2 Volt Mid Value – This field displays the expected midpoint voltage input
value in a three point calibration provided by the user during calibration.
#3 Volt Max Value – This field displays the expected maximum voltage input
value provided by the user during calibration.
#1 Raw Min Value – This field displays the measured minimum value read
from the device.
#2 Raw Mid Value – This field displays the measured midpoint value read
from the device.
#3 Raw Max Value – This field displays the measured maximum value read
from the device.
Calibration Points – This field displays the number of points the current
information used to calibrate the voltage input.
Status – This field displays Live or Manual, depending on the operating mode
selected, as well as low and high alarm conditions.
Содержание DensityPRO NAI Series
Страница 2: ......
Страница 4: ......
Страница 6: ...This page intentionally left blank ...
Страница 8: ...This page intentionally left blank ...
Страница 14: ...xiv DensityPRO User Manual Thermo Scientific This page intentionally left blank ...
Страница 22: ...A xxii DensityPRO User Manual Thermo Scientific Quick Setup This page intentionally left blank ...
Страница 124: ...Operation Application Screens 4 72 DensityPRO User Manual Thermo Scientific Figure 4 39 High and Low Alarms ...
Страница 134: ......
Страница 158: ......
Страница 195: ......