Operation
Physical Inputs & Outputs
Thermo Scientific
DensityPRO User Manual
4-23
Voltage Tabs
To access the information on the Voltage Input tabs using only the remote transmitter
keypad, see
Figure 4-12.
Voltage Input Tabs
The voltage inputs can operate in Live mode or Manual mode. In the Manual mode,
the user may enter a value in the Manual Value (EU) textbox. In Live mode, the user
does not have this option.
Once a value in engineering units has been entered into the Manual Value (EU) field,
the EZ Cal II software will convert that value into volts and display that value in the
Present Value field.
1.
Choose the type of input from the selections in the dropdown list.
Not Used
Temperature (Used for Density)
Pressure (Used for Level)
Flow (Used for Density)
Density (Used for Level)
Based on the input type selected, the fields available for configuration and the
information contained in the dropdown lists may change. See
for units available based on input selection.
2.
Define the input operating range by entering the Minimum Volt and Maximum
Volt into the corresponding textboxes.
Содержание DensityPRO NAI Series
Страница 2: ......
Страница 4: ......
Страница 6: ...This page intentionally left blank ...
Страница 8: ...This page intentionally left blank ...
Страница 14: ...xiv DensityPRO User Manual Thermo Scientific This page intentionally left blank ...
Страница 22: ...A xxii DensityPRO User Manual Thermo Scientific Quick Setup This page intentionally left blank ...
Страница 124: ...Operation Application Screens 4 72 DensityPRO User Manual Thermo Scientific Figure 4 39 High and Low Alarms ...
Страница 134: ......
Страница 158: ......
Страница 195: ......