Servicing
Pump Replacement
Thermo Scientific
42iQ Instruction Manual
7-19
Use the following procedure to replace the pump.
Equipment required:
Phillips drive, #1 and #2
1.
Turn instrument OFF, unplug power cord, and remove the cover
(
Figure 2–1
).
2.
Unplug pump cable from Step POL board J7.
3.
Twist opposite direction to unlock tube clamps.
Note
Push in tube clamp to lock.
▲
4.
Disconnect tubing from pump.
5.
Unfasten two captive hardware.
6.
Slide pump left until keyway meets opening.
Pump
Replacement