Optional Equipment
Internal Permeation Span Source
Thermo Scientific
42iQ Instruction Manual
9-23
5.
Pull thermistor assembly upwards.
Figure 9–13.
Installing Glass Tube pt 2
6.
Push and gently twist tube dispersion glass upwards and seat into o-
ring.
Note
Make sure PTFE tube and thermistor is inside the dispersion glass.
Keep glass clean by using kimwipes or similar material to handle glass.
▲