Optional Equipment
Internal Permeation Span Source
9-22
42iQ Instruction Manual
Thermo Scientific
Use the following procedure to install the permeation tube.
CAUTION
Allow oven to cool down prior to servicing.
▲
1.
Lift oven cover up.
2.
Lift and unlatch two oven cover handles.
Figure 9–12.
Installing Glass Tube pt 1
3.
Unfasten nut. Loosen thumb screw 2X full counter clock rotation.
4.
Loosening the thumb screw deflates the o-ring making it easier to pull
the thermistor assembly out.
Permeation Tube
Installation