Servicing
LCD Module Replacement
7-12
42iQ Instruction Manual
Thermo Scientific
Use the following procedure to replace the LCD module.
Equipment required:
Wrench, 1/4
1.
Turn instrument OFF and unplug the power cord.
2.
Gripping from the top corners of the front panel and pull outwards.
3.
Unfasten four nuts (
Figure 7–13
).
Figure 7–13.
Replacing the LCD Module
4.
Remove cover.
5.
Unplug LCD cables from backside of board.
6.
Pull board off the standoffs.
LCD Module
Replacement