Operation
Calibration
Thermo Scientific
42iQ Instruction Manual
3-15
The Calibrate NO, NO
2
, and NO
x
Span Coefficient screens are used to
enter span concentrations and calibrate the NO, NO
2
, and NO
x
span
coefficients. The NO span coefficient is calculated, stored, and used to
correct the current reading. All calibration screens function the same way.
Therefore, the following example of the NO screen applies to the NO
2
and
NO
x
calibration screens as well.
The following screens are shown in single range mode and dual or auto
range mode. In dual or auto range modes, “High” or “Low” is displayed to
indicate the calibration of the high or low coefficient. The Calibrate High
Span Coefficient and Calibrate Low Span Coefficient screens function the
same way.
It is important to note the averaging time when calibrating. The longer the
averaging time the more precise the calibration results. To achieve
maximum precision, allow the instrument to stabilize each time input gas is
changed and set the averaging time to 300-second averaging.
Home Screen>Calibration>Calibrate Span Coefficient (single range
mode)
Calibrate NO, NO
2
, and
NO
x
Span Coefficient
Содержание 42iQ
Страница 1: ...42iQ Instruction Manual NO NO2 NOx Analyzer 117435 00 1Apr2020 ...
Страница 2: ......
Страница 18: ......
Страница 34: ...Operation Instrument Display 3 8 42iQ Instruction Manual Thermo Scientific ...
Страница 164: ......
Страница 192: ......
Страница 199: ...Maintenance Pump Rebuilding Thermo Scientific 42iQ Instruction Manual 5 7 Figure 5 3 Pump Rebuilding ...
Страница 214: ...Servicing Fan Replacement 7 8 42iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 256: ...Servicing DMC Ozonator 7 50 42iQ Instruction Manual Thermo Scientific Figure 7 53 Remove Ammonia Scrubber ...
Страница 306: ......
Страница 318: ......
Страница 319: ......