Operation
Settings
Thermo Scientific
42iQ Instruction Manual
3-85
The Advanced Measurement Settings screen allows the user to calibrate the
reaction chamber and set other advanced settings.
Home Screen>Settings>Measurement Settings>Advanced Measurement
Settings
The Advanced Measurements Settings screen contains the following
information:
●
Reaction Chamber Settings:
User can calibrate Input board and PMT.
●
Perm Oven Selection:
Allows the user to select the operating temperature
of the perm oven.
●
Auto/Manual Mode:
User can choose to measure single gas or switching
mode.
●
Extended Ranges:
Enables/disables extended ranges feature.
●
Compensation:
Allows the user to compensate for changes in
temperature and pressure concentration.
●
Pressure Calibration:
Calibrates pressure.
Advanced Measurement
Settings