Calibration
Equipment Required
Thermo Scientific
42iQ Instruction Manual
4-3
Fixed bed reactors are commonly used in the last step of zero air generation
to remove the remaining contaminants by either further reaction or
absorption.
Table 4–1
lists materials that can be effective in removing
contaminants.
Table 4–1.
Scrubbing Materials
To Remove
Use
NO
Soda-Lime (6-12 mesh), Purafil
Hydrocarbons
Molecular Sieve (4A), Activated Charcoal
O
3
and NO
2
Activated
Charcoal
A gas phase titrator (GPT), such as is included in the Thermo Scientific
146iQ Multi-gas Calibrator
,
is used to generate NO
2
concentrations from
NO concentrations.
Figure 4–1
shows the suggested placement of the
component parts of a gas phase titration apparatus.
Equipment Damage
All connections between components in the system
should be made with glass, PTFE, or other non-reactive material.
▲
The airflow controllers should be devices capable of maintaining constant
airflows within
2% of the required flow rate. The NO flow controller
should be capable of maintaining constant NO flows within
2% of the
required flow rate.
The pressure regulator for the standard NO cylinder must have a non-
reactive diaphragm and internal parts, and a suitable delivery pressure.
Scrubbing
Gas Phase Titrator
Flow Controllers
Pressure Regulator
Содержание 42iQ
Страница 1: ...42iQ Instruction Manual NO NO2 NOx Analyzer 117435 00 1Apr2020 ...
Страница 2: ......
Страница 18: ......
Страница 34: ...Operation Instrument Display 3 8 42iQ Instruction Manual Thermo Scientific ...
Страница 164: ......
Страница 192: ......
Страница 199: ...Maintenance Pump Rebuilding Thermo Scientific 42iQ Instruction Manual 5 7 Figure 5 3 Pump Rebuilding ...
Страница 214: ...Servicing Fan Replacement 7 8 42iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 256: ...Servicing DMC Ozonator 7 50 42iQ Instruction Manual Thermo Scientific Figure 7 53 Remove Ammonia Scrubber ...
Страница 306: ......
Страница 318: ......
Страница 319: ......