Servicing
DMC Converter Replacement
7-42
42iQ Instruction Manual
Thermo Scientific
Use the following procedure to replace the converter cartridge heater.
Equipment Required:
Phillips drive, #2
Hex drive, 1/4-inch
1.
Turn the instrument OFF, unplug the power cord, and remove the
cover.
2.
Remove DMC Converter as described in “
DMC Converter
Replacement
” in this chapter.
3.
Using a 1/4-inch hex drive, unfasten six 6-32 hex head screws.
4.
Take apart top housing assembly and insulator to get to the heater
assembly.
5.
Remove the converter cartridge/heater assembly from the bottom
housing assembly.
6.
Remove fitting body to allow clearance for heater to slide out.
7.
Loosen the heater clamp, pry heater apart no wider than necessary and
remove the converter cartridge noting the proper orientation of heater
wires and thermocouple probe.
8.
To replace converter, follow previous steps in reverse. Note be sure to
wrap the O
3
converter tube snugly around the heater.
9.
Reinstall in reverse order.
Converter Cartridge
Heater Replacement
Содержание 42iQ
Страница 1: ...42iQ Instruction Manual NO NO2 NOx Analyzer 117435 00 1Apr2020 ...
Страница 2: ......
Страница 18: ......
Страница 34: ...Operation Instrument Display 3 8 42iQ Instruction Manual Thermo Scientific ...
Страница 164: ......
Страница 192: ......
Страница 199: ...Maintenance Pump Rebuilding Thermo Scientific 42iQ Instruction Manual 5 7 Figure 5 3 Pump Rebuilding ...
Страница 214: ...Servicing Fan Replacement 7 8 42iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 256: ...Servicing DMC Ozonator 7 50 42iQ Instruction Manual Thermo Scientific Figure 7 53 Remove Ammonia Scrubber ...
Страница 306: ......
Страница 318: ......
Страница 319: ......