Servicing
Optional Solenoid Valves Removal
Thermo Scientific
42iQ Instruction Manual
7-53
Use the following procedure to replace the mode valve, and optional
permeation oven valve and lag valve.
Equipment Required:
Phillips drive, #1
CAUTION
Allow the oven to cool down prior to servicing.
▲
1.
Turn the instrument OFF, unplug the power cord, and remove the
cover (
Figure 2–1
).
2.
Unplug mode valve (J5), lag valve (J6), and perm valve (J7) from step
board 2 only.
Figure 7–57.
Mode-Lag-Perm Electrical Connection
Optional
Solenoid Valves
Removal