Installation and Setup
Cover Removing and Replacing
2-2
42iQ Instruction Manual
Thermo Scientific
Use the following procedure to remove and replace the cover.
Equipment required:
Phillips screwdriver, #2
1.
Unfasten the four 8-32 screws securing the cover (shipping screws).
2.
Press in both latches located on top cover and hold while pulling up to
remove. Set upright.
Figure 2–1.
Removing the Cover
3.
To replace, align cover and drop in. Latches will automatically snap in
place.
Cover Removing
and Replacing