Servicing
I/O Replacement
7-14
42iQ Instruction Manual
Thermo Scientific
Use the following procedure to replace the I/O boards.
8.
Turn instrument OFF, unplug power cord, and remove the cover
(
Figure 2–1
).
9.
Swing arm open.
Figure 7–15.
I/O Replacement, Arm
10.
Pull board upwards.
Figure 7–16.
I/O Replacement, Remove Board
I/O Replacement