Calibration
Equipment Required
Thermo Scientific
42iQ Instruction Manual
4-7
The O
3
generator (ozonator) airflow rate and the NO flow rate must be
adjusted such that the following relationships hold:
min
-
ppm
2.75
t
x
]
[NO
=
P
R
RC
R
)
F
+
F
(
F
]
[NO
=
]
[NO
NO
O
NO
STD
RC
min
2
<
F
+
F
V
=
t
NO
O
RC
R
Use the following procedure to determine the flow conditions to be used in
the GPT system.
1.
Determine FT, the total flow required at the output manifold, which
should be equal to the analyzer demand plus 10 to 50 percent excess.
2.
Establish [NO]
OUT
as the highest NO concentration that will be
required at the output manifold. [NO]
OUT
should be about equal to
90% of the upper range limit (URL) of the NO
2
concentration range to
be covered.
3.
Determine F
NO
as:
]
[NO
F
x
]
[NO
=
F
STD
T
OUT
NO
4.
Select a convenient or available reaction chamber volume. Initially a
trial volume may be selected in the range of 200 to 500 cc.
5.
Compute F
O
as:
F
-
2.75
V
x
F
x
]
[NO
=
F
NO
RC
NO
STD
O
6.
Compute t
R
as:
Determining GPT System
Flow Conditions
Содержание 42iQ
Страница 1: ...42iQ Instruction Manual NO NO2 NOx Analyzer 117435 00 1Apr2020 ...
Страница 2: ......
Страница 18: ......
Страница 34: ...Operation Instrument Display 3 8 42iQ Instruction Manual Thermo Scientific ...
Страница 164: ......
Страница 192: ......
Страница 199: ...Maintenance Pump Rebuilding Thermo Scientific 42iQ Instruction Manual 5 7 Figure 5 3 Pump Rebuilding ...
Страница 214: ...Servicing Fan Replacement 7 8 42iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 256: ...Servicing DMC Ozonator 7 50 42iQ Instruction Manual Thermo Scientific Figure 7 53 Remove Ammonia Scrubber ...
Страница 306: ......
Страница 318: ......
Страница 319: ......