Operation
Settings
Thermo Scientific
42iQ Instruction Manual
3-81
●
High Range NO Setting:
Sets the high range NO concentration range
for the analog output when in dual or auto range mode.
●
Low Range NO Setting:
Sets the low range NO concentration range for
the analog output when in dual or auto range mode.
●
NO
2
Range Setting:
Sets the NO
2
concentration range for the analog
outputs when in single range mode.
●
High Range NO
2
Setting:
Sets the high range NO
2
concentration range
for the analog output when in dual or auto range mode.
●
Low Range NO
2
Setting:
Sets the low range NO
2
concentration range
for the analog output when in dual or auto range mode.
●
NO
x
Range Setting:
Sets the NO
x
concentration range for the analog
outputs when in single range mode.
●
High Range NO
x
Setting:
Sets the high range NO
x
concentration range
for the analog output when in dual or auto range mode.
●
Low Range NO
x
Setting:
Sets the low range NO
x
concentration range
for the analog output when in dual or auto range mode.
Settable standard ranges according to unit selection include:
ppb 50
20000 ppb
ppm 0.05
20 ppm
% 0.000005
0.002 %
μg/m
3
100
30000 μg/m
3
mg/m
3
0.1
30 mg/m
3
g/m
3
.0001
0.030 g/m
3
Settable extended ranges according to unit selection include:
ppb 200
100000 ppb
ppm 0.2
100 ppm
% 0.00002
0.01 %
Содержание 42iQ
Страница 1: ...42iQ Instruction Manual NO NO2 NOx Analyzer 117435 00 1Apr2020 ...
Страница 2: ......
Страница 18: ......
Страница 34: ...Operation Instrument Display 3 8 42iQ Instruction Manual Thermo Scientific ...
Страница 164: ......
Страница 192: ......
Страница 199: ...Maintenance Pump Rebuilding Thermo Scientific 42iQ Instruction Manual 5 7 Figure 5 3 Pump Rebuilding ...
Страница 214: ...Servicing Fan Replacement 7 8 42iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 256: ...Servicing DMC Ozonator 7 50 42iQ Instruction Manual Thermo Scientific Figure 7 53 Remove Ammonia Scrubber ...
Страница 306: ......
Страница 318: ......
Страница 319: ......