Sierra M6
‐
2 SAS/SATA Protocol Analyzer User Manual
191
Target and Device Emulation
Teledyne LeCroy
Frame Length Error
Check Frame Length Error and choose the type of error to introduce.
Under
frame
length
: Data frame payload returned by the DE is 0 bytes.
Set
data
frame
length
to
xxxx
DWORDs
: Sets the frame length and not the pay
‐
load length. You need to consider the frame header in the calculations.
Over
frame
length
: 2049 DWORD payloads are sent instead of 2048, because
four 00 bytes are added to the payload.
Code violation and Disparity errors
Check these and specify the DWORD offset and the number of error DWORDs.
Don’t check credit (SAS only)
Check this to disable credit checking.
Value of Reserved Fields
Enter a value.
Ignore XRDY/RRDY sequence (SATA only)
Check or not.
Delay in HOLD Response (SATA only)
Check or not.
No WTRM (SATA only)
Check or not.
SOF, EOF & CRC Errors
Check any or all and choose the criteria for introduction.
Insert Primitive
Check Insert Primitive, click the Type button to open the Primitive Type dialog, then select
the primitive type.
Send Primitive
Check
Send
stand
‐
alone
primitive
, click the Type button to open the Primitive Type
dialog, select the primitive type, and click
OK
.
Figure 2.204: Primitive Type
Содержание Sierra M6-2
Страница 37: ...Sierra M6 2 SAS SATA Protocol Analyzer User Manual 35 Expandability Teledyne LeCroy ...
Страница 375: ...Sierra M6 2 SAS SATA Protocol Analyzer User Manual 373 Interface Teledyne LeCroy ...
Страница 436: ...Teledyne LeCroy Scenario Batch Files 434 Sierra M6 2 SAS SATA Protocol Analyzer User Manual ...
Страница 526: ...Teledyne LeCroy 524 Sierra M6 2 SAS SATA Protocol Analyzer User Manual ...