ADFM® Flow Meter
Section 3 Maintenance and Troubleshooting
This section explains the maintenance requirements of the
ADFM Velocity Profiling System.
The ADFM is designed to perform reliably in adverse conditions
with a minimal amount of routine service requirements. To keep
your system working properly, inspect the battery voltage and
the condition of the internal desiccant regularly.
Maintenance intervals are affected by many variables. The Data
Storage Rate will affect the battery life. Humidity levels obvi-
ously affect the service life of the desiccant, and the amount of
debris in the stream can drastically alter the channel conditions.
3.1 How to Obtain Service
If you think your equipment requires repair, contact Teledyne
Isco’s Technical Service Department.
Teledyne Isco
Technical Service Dept.
P.O. Box 82531
Lincoln, NE 68501 USA
Phone: (866) 298-6174
(402) 464-0231
FAX: (402) 465-3085
E-mail:
[email protected]
The cable connections and electronic components of the sensor
are encapsulated in plastic resin and are not user-serviceable.
Corresponding with a Teledyne Isco Technical Service Represen-
tative can often resolve the problem without the need to return
the item. If the difficulty cannot be resolved you will be issued a
Return Authorization Number (RAN) and information for
returning it to the factory.
Before packaging and shipment, clean and dry the flow meter
and sensor, and include important information as described in
the following sections.
3.1.1 Cleaning
The electronics case, cables, and sensor may be cleaned with mild
detergent and warm water. Before cleaning, ensure that all pro-
tective connector caps are in place. Do not use abrasive agents or
solvents; they will damage the surfaces of the equipment. Mild
chlorine bleach solutions may be used for disinfection and elimi-
nation of odors.
3.1.2 Packaging
Attach documentation to the equipment identifying the owner
and owner address, complete instrument model and serial
number, and a description of the problems observed with the
equipment.
Whenever possible, the original factory packaging material
should be retained for reuse. For repackaging with commercial
materials, follow these instructions:
• Wrap the flow meter and sensor assemblies separately
in bubble wrap or other cushioning material.
• Use a strong shipping container suitable for the weight
of the ADFM. Shipping containers made of wood or
Содержание ADFM Hot Tap
Страница 2: ......
Страница 4: ......
Страница 20: ...ADFM Flow Meter Section 1 Introduction 1 10 ...
Страница 85: ...ADFM Flow Meter Section 2 Setup and Installation 2 65 Figure 2 73 Viewing data in Expert View ...
Страница 86: ...ADFM Flow Meter Section 2 Setup and Installation 2 66 ...
Страница 98: ...ADFM Flow Meter Section 3 Maintenance and Troubleshooting 3 12 ...
Страница 114: ......
Страница 116: ......