ADFM® Flow Meter
Section 2 Setup and Installation
2-23
Console Enclosure
Mounting
Stainless steel environmental console enclosures are available
from Teledyne Isco for outdoor or indoor installations without
climate control, and where dust or moisture is present. The elec-
tronics box and all other electronics must be mounted within the
enclosure.
The environmental console enclosure can be ordered with or
without a 400- or 800-watt heater; contact Teledyne Isco for more
information.
Where a console enclosure will be used, a user-supplied concrete
pad and steel stand, or other secure method, as well as cable
conduit with minimum 2" ID, are recommended.
Note
Environmental enclosures are not shipped with access holes
for conduit connection.
All cable entries must have appropriate ID, user-supplied con-
duit connections or cord-grip fittings to retain the NEMA 4X rat-
ing. Conduit and wires must be sealed to prevent harmful
gases and moisture from entering the enclosure.
Failure to
seal conduit could reduce equipment life.
Figure 2-20 ADFM controller: environmental console enclosure