ADFM® Flow Meter
Section 2 Setup and Installation
2-49
Figure 2-51 Align the edges of the box parallel to flow
5. Loosen the two Phillips screws on the back of the two-piece
clamping collar.
WARNING
Never loosen any hardware on the Hot Tap sensor when the
pipe is pressurized. Pressure within the pipe can force the
shaft to come out of the pipe, causing grave injury to
anyone nearby.
6. Lock the sensor shaft in place by evenly tightening the two
hex socket screws in the sides of the two-piece clamping
collar, and then re-tighten the two Phillips screws.
Safety reinforcement
The sensor assembly includes a tethering chain as a precau-
tionary measure in the event that the clamping collar is inadver-
tently loosened.
Attach the quick disconnect links through holes in each of the
two flanges on the sensor shaft, adjusting the chain to proper
length.
Figure 2-52 Attaching the tether chain
Flow
WARNING