Appendix D: Waveform Transfer (WFMOutpre and CURVe Query) Examples
WFMOutpre? Query results
Explanation
ENCDG ASCII
This value speci
fi
es the encoding of the waveform data. To change this value (the
other possibility is BINARY), use the
command. (This value can
also be set using the
command, which provides the ability to set the
WFMOutpre:ENCdg, WFMOutpre:BN_Fmt,
: and
WFMOutpre:BYT_Or
values
using a single command.)
BN_FMT RI
This value speci
fi
es the binary format, which in this case is RI (signed integer). To change
this value (the other possibility is RP or positive integer), use the
command. Note: this
fi
eld is not applicable for ASCii encoding.
BYT_OR MSB
This value speci
fi
es the byte order for the BINARY encoding, which in this case is MSB
(most signi
fi
cant byte
fi
rst, also known as IBM format). To change this value to LSB, use the
command. Note: this
fi
eld is not applicable for ASCii encoding.
WFID "Digital, unknown
coupling, 100.0us/div,
10000 points, Digital
mode"
This quoted string represents information about the source waveform that would be returned
by a.
WFMOutpre:WFId?
query. It cannot be changed.
NR_PT 25
This value indicates the number of data points in the waveform record to be transferred
using the CURVE? query. (If you would like to determine only this value, use the
query.) Note: this value is typically equal to the full record length,
but you also have the option to transfer only a portion of the record length by using the
and
commands.
PT_FMT Y
This value indicates the format of the data points in the waveform record. In this case,
the value represents YT format. This is query only – the returned values can be Y for YT
format or ENV for envelope format (min/max pairs). (If you would like to determine only this
value, use the
query.)
PT_ORDER LINEAR
This value is always
LINear
.
XUNIT "s"
This value indicates the units of the x-axis of the waveform record. This is determined by the
horizontal settings for the waveform source. For live channels, this value is “s”, representing
seconds. When using the math waveform as a source, the value can be “s” or “Hz”. This
is query only. (If you would like to determine only this value, use the
query.)
XINCR 100.0000E-9
This value indicates the time, in seconds, or frequency, in hertz, between data points in the
waveform record. This is query only. (If you would like to determine only this value, use
the
query.)
XZER0 -500.0000E-6
This value indicates the time, in seconds, or frequency, in hertz, of the
fi
rst data point in the
waveform record. This time or frequency is relative to the time of the trigger, which is always
0. So, this XZEro time or frequency can be negative. This is query only. (If you would like to
determine only this value, use the
query.)
PT_OFF 0
This is a query provided only for compatibility with performance oscilloscopes. The
returned value is always 0. (If you would like to determine only this value, use the
query.)
YUNIT "State"
This value indicates the vertical units of data points in the waveform record. This can be
any of several string values, depending upon the vertical units of the source waveform –
in this case, State. This is query only. (If you would like to determine only this value, use
the
query.)
MSO54, MSO56, MSO58, MSO58LP Programmer
D-7
Содержание MSO54
Страница 1: ...xx 5 Series MSO MSO54 MSO56 MSO58 MSO58LP ZZZ Programmer Manual P077130502 077 1305 02 ...
Страница 2: ......
Страница 3: ...5 Series MSO MSO54 MSO56 MSO58 MSO58LP ZZZ Programmer Manual xx www tek com 077 1305 02 ...
Страница 8: ...Preface iv MSO54 MSO56 MSO58 MSO58LP Programmer ...
Страница 26: ...Command Syntax 2 10 MSO54 MSO56 MSO58 MSO58LP Programmer ...
Страница 936: ...Status and Events 3 20 MSO54 MSO56 MSO58 MSO58LP Programmer ...
Страница 937: ...Appendix A Character Set MSO54 MSO56 MSO58 MSO58LP Programmer A 1 ...
Страница 938: ...Appendix A Character Set A 2 MSO54 MSO56 MSO58 MSO58LP Programmer ...
Страница 996: ...Appendix D Waveform Transfer WFMOutpre and CURVe Query Examples D 18 MSO54 MSO56 MSO58 MSO58LP Programmer ...