Commands listed in alphabetical order
POWer:POWer<x>:SOA:ISOURce
This command sets or queries the current source for SOA measurement in the
speci
fi
ed power measurement badge. The power measurement badge is speci
fi
ed
by x.
Conditions
Requires option 5-PWR, SUP5-PWR, or 5-PS2.
Group
Power
Syntax
POWer:POWer<x>:SOA:ISOURce
{CH<x> | MATH<x> | REF<x>}
POWer:POWer<x>:SOA:ISOURce?
Arguments
CH<x>
= A channel speci
fi
er; <x> is 1 through 8 and is limited by the number
of FlexChannels in your instrument.
MATH<x>
= A math waveform speci
fi
er; <x> is
≥
1.
REF<x>
= A reference waveform speci
fi
er; <x> is
≥
1.
Examples
POWer:POWer1:SOA:ISOURce CH2
sets the current source for SOA
measurement as channel 2.
POWer:POWer<x>:SOA:POINT
This command sets or queries the Y co-ordinate value for SOA mask.
Conditions
Requires option 5-PWR, SUP5-PWR, or 5-PS2.
Group
Power
Syntax
POWer:POWer<x>:SOA:POINT
<QString> <NR1>
POWer:POWer<x>:SOA:POINT
<QString>?
Arguments
<QString>
= 1Y | 2Y | 3Y | 4Y | 5Y | 6Y | 7Y | 8Y | 9Y | 10Y | 11Y | 12Y | 13Y
| 14Y | 15Y | 16Y | 17Y | 18Y | 19Y | 20Y | 21Y | 22Y | 23Y | 24Y | 25Y | 26Y
| 27Y | 28Y | 29Y | 30Y | 31Y | 32Y
<NR1>
ranges from –5000 to 5000
2-582
MSO54, MSO56, MSO58, MSO58LP Programmer
Содержание MSO54
Страница 1: ...xx 5 Series MSO MSO54 MSO56 MSO58 MSO58LP ZZZ Programmer Manual P077130502 077 1305 02 ...
Страница 2: ......
Страница 3: ...5 Series MSO MSO54 MSO56 MSO58 MSO58LP ZZZ Programmer Manual xx www tek com 077 1305 02 ...
Страница 8: ...Preface iv MSO54 MSO56 MSO58 MSO58LP Programmer ...
Страница 26: ...Command Syntax 2 10 MSO54 MSO56 MSO58 MSO58LP Programmer ...
Страница 936: ...Status and Events 3 20 MSO54 MSO56 MSO58 MSO58LP Programmer ...
Страница 937: ...Appendix A Character Set MSO54 MSO56 MSO58 MSO58LP Programmer A 1 ...
Страница 938: ...Appendix A Character Set A 2 MSO54 MSO56 MSO58 MSO58LP Programmer ...
Страница 996: ...Appendix D Waveform Transfer WFMOutpre and CURVe Query Examples D 18 MSO54 MSO56 MSO58 MSO58LP Programmer ...